Product parameters:
Instrument model: WXJ-III
Measurement range: CODcr: 5-2500 mg/L;
TP:0.01~0.6mg/L;
TN:0.05~4mg/L;
CODcr's ability to resist chloride ion interference: ≤ 10000 mg/L
Precision: When the CODcr value is between 80-150mg/L, the indoor relative standard deviation is ≤ 4.3%, and the inter room relative standard deviation is ≤ 5.0%
TP and TN can meet the requirements of the national standard (GB11889-11915-89) for water quality vocabulary (Part 3-7) and analytical methods
Accuracy: CODcr compared to classical reflux method, correlation coefficient r>0.999
Comparison of TP, TN and standard methods, correlation coefficient r>0.999
Sample quantity for digestion (per batch): Standard configuration includes 6 samples
Dissolving time: 5-16 minutes
Microwave operating frequency: 2450MHz
Rated input power: 1300W (maximum)
Microwave output power: 800W
Dimensions: 483x389x281mm
- Product Details
- Product manual
Yetuo WXJ-III Microwave Digestion Instrument
Model: WXJ-III
Product Overview:
Widely applicable to the identification and management of water quality by environmental protection departments, water resources management departments, and public health departments at all levels. It can quickly and efficiently determine the chemical oxygen demand (CODCr), total phosphorus (TP), and total nitrogen (TN) in various surface water, domestic sewage, and industrial wastewater.
Product Features of Yetuo WXJ-III Microwave Digestion Instrument
Sealed digestion, overpressure protection, safe and reliable, with a buzzer indicating completion of digestion;
In CODcr analysis, it effectively inhibits the oxidation of chloride ions to chlorine gas by potassium dichromate;
When dissolving CODcr, it can resist interference from chloride ions;
Microwave energy heating method, with molecules rotating at high speed in the digestion system, resulting in rapid heating and low consumption;
There are ventilation openings at the bottom and top of the device for heat dissipation and exhaust;
● Note: The digestion tank is made of polytetrafluoroethylene and is implemented according to the specified reagents and reaction conditions. It cannot be used for other purposes